CCR 365, 0.1 Torr F.S.
  • Order number: PT R27 643
  • Measurement range from 1 · 10<sup>-5</sup> – 0.13 hPa (1 · 10<sup>-5</sup> – 0.1 Torr)
  • Accuracy: 0.5 % of measurement
  • Supply voltage: 14 – 30 V
  • Output signal: 0 – 10 V
  • Ceramic technology sensor
  • No memory effects
  • Precise temperature compensation
  • Excellent zero stability
  • Output signal an connector compatible with MKS Baratron
  • Calibration test report included in delivery
  • Delivery time: On request
€1,860.00
Connection flange ½" VCR
Resolution 0.003 % F.S.
Output signal: Measuring range 0 – 10 V
Output signal: Minimum load 10 kΩ
Bakeout temperature at the flange ≤ 110 °C
Pressure max. 1300 hPa
Input voltage(s) 14 – 30 V DC
Flange, material Stainless steel 1.4404 (AISI 316L)
Full scale 0.133 hPa
Weight 0.37 kg
Power consumption max. 1 W
Materials in contact with media
  • Ceramic (Al<sub>2</sub>O<sub>3</sub> = 99.5 %)
  • Stainless steel (AISI 316L)
Measuring range 1.33E-5 – 1.33E-1 hPa
Accuracy of measurement in range 1 0.5 % (of measured value)
Measurement cable length 100 m
Measuring method
  • Capacitive
Response time 130 ms
Protection degree
  • IP30
Temperature: Storage -40 – 65 °C
Temperature effect: on range 0.03 % of measured value/ °C
Temperature effect: on zero point 0.02 % F.S./°C
Ambient temperature 5 – 50 °C
Volume ≤ 3.6 cm³
Dimension Image