CCT 361, 8-VCR
  • Order number: PT R50 435
  • Measurement range: 0.1 – 1100 hPa
  • Pressure measurement independent of type of gas
  • Minimal zero drift
  • Precise temperature compensation
  • Sensor in ceramic technology
  • Calibration test report included in delivery
  • Profinet interface
  • Delivery time: 6 weeks
€2,100.00
Connection flange ½" VCR
Resolution 0.003 % F.S.
Bakeout temperature at the flange ≤ 110 °C
Pressure max. 4000 hPa
I/O interfaces RS-485, Profinet
Input voltage(s) 14 – 30 V DC
Electrical connection RS485, M12, 5-pole, socket, a-coded
Full scale 1000 hPa
Weight 0.7 g
Power consumption max. 5 W
Materials in contact with media
  • Ceramic (Al<sub>2</sub>O<sub>3</sub> = 99.5 %)
  • Stainless steel (AISI 316L)
Measuring range 1E-1 – 1.1E3 hPa
Accuracy of measurement in range 1 0.2 % (of measured value)
Measurement cable length 100 m
Measuring method
  • Capacitive
Measuring cycle 60 ms
Interface: Connection, device side Profinet: 2 x M12, 5-pole, socket, d-coded
Protection degree
  • IP40
Temperature: Storage -40 – 65 °C
Temperature effect: on range 0.01% of measured value/ °C
Temperature effect: on zero point 0,005 % F.S./°C
Ambient temperature 5 – 50 °C
Dimension Image