TPR 280, 250 °C
  • Order number: PT R26 960
  • Flange size DN 16 ISO-KF
  • Maximum pressure refers to inert gases
  • Compact and rugged
  • With extra long housing
  • Delivery time: 6 weeks
€480.00
Connection flange DN 16 ISO-KF
Resolution 1 % of reading
Output signal: Measuring range 1.5 – 8.5 V
Output signal: Minimum load 10 kΩ
Bakeout temperature at the flange ≤ 250 °C
Seal
  • Metal
Pressure max. 10000 hPa
Feedthrough
  • Glass
Attribute Stainless steel, metal sealed
Input voltage(s) 14 – 30 V DC
Flange, material Stainless steel
Weight 0.13 kg
Filament Tungsten
Power consumption max. 1 W
Measuring range 1E-4 – 1E3 hPa
Accuracy of measurement in range 1 15 % (1E-3 hPa – 1E2 hPa)
Measurement cable length 200 m
Measuring method
  • Pirani
Response time 80 ms
Protection degree
  • IP40
Temperature: Storage -20 – 65 °C
Ambient temperature 5 – 60 °C
Volume 10 cm³
Repeatability in range 1 2 % (1E-3 hPa – 1E2 hPa)
Dimension Image