IKR 270, metal-sealed
  • Order number: PT R21 251
  • Flange size DN 40 CF-F
  • The maximum pressure applies to inert gases and temperatures below 55 °C
  • Cold cathode (inverted magnetron)
  • Rugged and dependable
  • Insensitive to air ingress
  • Corrosion resistant
  • Delivery time: 1-3 days
€1,510.00
Anode
  • Mo
Connection flange DN 40 CF-F
Output signal: Measuring range 1.96 – 8.6 V
Output signal: Minimum load 10 kΩ
Bakeout temperature max.(electronics removed) ≤ 250 °C
Seal
  • Silver
Pressure max. 1000 kPa
Feedthrough
  • Al<sub>2</sub>O<sub>3</sub>
Attribute Metal sealed
Input voltage(s) 15 – 30 V DC
Flange, material Stainless steel
Weight 0.95 kg
Power consumption max. 2 W
Measuring range 5E-11 – 1E-2 hPa
Accuracy of measurement in range 1 30 % (1E-9 hPa – 1E-3 hPa)
Measurement cable length 500 m
Measuring method
  • Cold cathode
Temperature: Storage -40 – 65 °C
Ambient temperature 5 – 55 °C
Volume 20 cm³
Repeatability in range 1 5 % (1E-9 hPa – 1E-3 hPa)
Dimension Image