CCR 375, 0,1 Torr F.S.
  • Order number: PT R28 140
  • Measurement range from 1.33 · 10-5 – 0.13 hPa
  • Outstanding long-term and temperature stability
  • Minimal zero drift
  • Corrosion-resistant ceramic technology
  • Sensor-Shield provides additional protection against contamination
  • Output signal an connector compatible with MKS Baratron
  • Calibration test report included in delivery
  • Delivery time: On request
€2,900.00
Resolution0.003 % F.S.
Output signal: Measuring range0 – 10 V
Output signal: Minimum load10 kΩ
Bakeout temperature at the flange110 °C
Pressure max.1300 hPa
Attribute2 switch-points
Full scale0.133 hPa
Precision: % of measurement0.15 %
Weight900 g
Materials in contact with mediaCeramics (Al2O3 = 99.5 %)
Diaphragm and measuring chamberCeramics (Al2O3 = 99.5 %)
Measuring range1.33E-5 – 1.33E-1 hPa
Measuring methodCapacitance
Nominal diameterTube OD ½"
Response time130 ms
Pipe and flangeStainless steel
Switch-point0 – 99 % F. S.
Number of switch-points2 Pcs.
Switch-point: Voltage DC30 V DC
Switch-point: Current DC0.5 A
Protection degreeIP40
Temperature: Storage-40 – 65 °C
Temperature effect: on range0.01% of measured value/ °C
Temperature effect: on zero point0,005 % F.S./°C
Ambient temperature10 – 40 °C
Supply: Voltage V DC14 – 30 V DC
Supply: consumption max.≤ 12 W
Volume≤ 4.2 cm³